4.10 Setting the Averaging
────────────────────────────────────────────────────
(N-1) A
n-1
+ Zn
Display value =
―――――――
N
Zn: nth measured data
A
n-1
: n-1 th display value
N: constant setting
F1
F2
F3
F4
EXPONENTIAL
TIME
NOTE
(3) Setting the Exponential Average (AV-E)
The exponential average provides a average of the previous values, but
weighted toward the latest value. The effect of previous values thus
diminishes exponentially.
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
TIME
" (time control) page.
2. Using the
CURSOR
keys, move the cursor to the
"
AVERAGING
" item.
3. Press
F3
(
Exponential
).
4. Move the cursor to the specified number of
sampling item on the right, and set the constant
value using the function key.
5. Press the
MEAS
key to return to the measurement
screen, and averaging starts.
This has no connection with any time settings.