Safety Notes
────────────────────────────────────────────────────
CATⅡ
Primary electrical circuits in equipment connected to an AC electrical outlet by a
power cord (portable tools, household appliances, etc.)
CAT II covers directly measuring electrical outlet receptacles.
CATⅢ
Primary electrical circuits of heavy equipment (fixed installations) connected
directly to the distribution panel, and feeders from the distribution panel to outlets.
CATⅣ
The circuit from the service drop to the service entrance, and to the power meter
and primary overcurrent protection device (distribution panel).
Measurement categories
9600, 9601 and 9602 instrument comply with CAT III (600 V or less)/ CAT
II (600 to 1000 V) safety requirements.
To ensure safe operation of measurement instruments, IEC 61010 establishes
safety standards for various electrical environments, categorized as CAT
Ⅱ
to CAT IV, and called measurement categories.
Using a measurement instrument in an environment designated with a
higher-numbered category than that for which the instrument is rated could
result in a severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT
IV measurement applications could result in a severe accident, and must be
carefully avoided.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......