249
────────────────────────────────────────────────────
16.3 Specifications (using with the HIOKI 3193)
────────────────────────────────────────────────────
16
17
18
19
5
6
7
8
9
10
11
12
13
14
A
Voltage (
U
)
Current (
I
)
Active power (
P
)
Input type
Resistor voltage divider
+ isolation amplifier,
for isolated input
Isolated input, using
current sensor
――――
Measurement type
Analog processing:
・
True effective value
measurement
・
Mean value
rectification effective
value measurement
・
DC measurement
Analog processing:
・
True effective value
measurement
・
Mean value rectification
effective value
measurement
・
DC measurement
Analog processing:
Active power
measurement
Input resistance
2 M
Ω±
100 k
Ω
(
±
5
%
) 200 k
Ω±
10 k
Ω
(
±
5
%
)
――――
Measurement range
6.0000/15.000/30.000
60.000/150.00/300.00
600.00 V
When using 20 A rated
sensor: 500.00 mA/
1.0000/ 2.0000/ 5.0000
10.000/ 20.000 A
When using 50 A rated
sensor: 1.2500/ 2.5000/
5.0000/ 12.500/ 25.000/
50.000A
When using 200 A rated
sensor: 5.0000/ 10.000/
20.000/ 50.000/ 100.00/
200.00 A
When using 500 A rated
sensor: 10.000/ 20.000/
50.000/ 100.00/ 200.00/
500.00 A
When using 1000 A rated
sensor:20.000/40.000/
100.00/200.00/400.00/
1000.0 A
(CT ratio = 2, the internal
operation range and
display range are the
same as the 500A sensor.)
Depending on
combination of voltage
and current values.
Effective input range
5% to 110% (up to 100% for 600 V range) or within the range accuracy
assurance range for the signal frequency specified elsewhere
Display range: 1% to 130% (when minimum range on current side, AC/
AC+DC) 0.2% to 130% (for others)
Maximum input voltage
Maximum input current
(55 Hz)
600 Vrms
850 Vpeak
Input range up to the
maximum input current of
the current sensor
――――
Crest factor
(Measurement range
×
6)
/measurement value or
Maximum input
voltage peak value/
measurement value
Lower of (measurement
range
×
6)/(measurement
value) and (input peak
value up to maximum
input current of the
current sensor)/
(measurement value)
――――
16.3 Specifications (using with the HIOKI 3193)
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......