132
────────────────────────────────────────────────────
*ESE0
────────────────────────────────────────────────────
Syntax
*ESE0
<NR1>
<NR1> = 0 to 255
Example
Transmission
*ESE0 34
Bits 5 and 1 of ESER0 are set to 34.
32+2=34
Syntax
*ESE0?
Response
syntax
Same as *ESE0
(Headers: ON) (Headers: OFF)
*ESE0
<NR1> <NR1>
Example
Transmission
Response
*ESE0?
*ESE0 34
34
Syntax
*ESE1
<NR1>
<NR1> = 0 to 255
Example
Transmission
*ESE1 74
Bits 6 and 3 of ESER1 are set to 74.
64+8=74
Function
Sets the mask pattern of the event status
enable register 0 (ESER0) to a value (0
to 255).
128
64
32
16
8
4
2
1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
SE
ST
PE
FE
SP
IE
CE
Note
When the power is turned on, and when a
reset has taken place upon key input, the
data is reinitialized to 0.
Error
Execution error / If the setting data is out
of the range.
Function
The contents of the event status enable
register 0 (ESER0) as set by the
*
ESE0
command are returned as a NR1 value (0
to 255).
Error
Query error / If the response message is
longer than 2000 bytes
Function
Sets the mask pattern of the event status
enable register 1 (ESER1) to a value (0
to 255).
128
64
32
16
8
4
2
1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
O6
O5
O4
O3
O2
O1
OF
Note
When the power is turned on, and when a
reset has taken place upon key input, the
data is reinitialized to 0.
Error
Execution error / If the setting data is out
of the range.
*ESE0
Sets the event status enable register 0.
*ESE0?
Queries the standard event status enable register 0 (SESER0).
*ESE1
Sets the event status enable register 1.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......