88
────────────────────────────────────────────────────
9.3 Internal Circuit for the External Control and Timing
────────────────────────────────────────────────────
CAUTION
・ To avoid damage to the unit, do not input voltage exceeding 5.5 V.
・ In the key lock state, external control functions are still enabled.
・ The input control signals should be clean signals, with no chattering.
External control terminal
Digital GND
9.3.1 INTEG.EXT.CONT and INTEG.RESET Terminals
5 V
(open-circuit)
0 V
(short-circuit)
5 V
(open-circuit)
0 V
(short-circuit)
Integration start
Integration stop
Reset during this time
300 ms
NOTE
9.3 Internal Circuit for the External Control and Timing
The external control signals can be 0/5 V logic signals or relay contact
open/closed circuit signals.
These terminals provide start/stop and reset control of integration, and have
the same function as the panel
START/STOP
key.
Start/stop and reset control of integration applies simultaneously to all items
being integrated. Separate control is not possible.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......