172
────────────────────────────────────────────────────
:MEASure:ITEM:EFFiciency
────────────────────────────────────────────────────
Syntax
:MEASure:ITEM:EFFiciency
<NR1>
<NR1> = 0 - 7
Example
:MEASURE:ITEM:EFFICIENCY 1
Sets the default items of the efficiency
measurement to EFF1.
Syntax
:MEASure:ITEM:EFFiciency?
Response
syntax
:MEASURE:ITEM:EFFICIENCY
<NR1>
Example
Transmission
Response
:MEASURE:ITEM:EFFICIENCY?
:MEASURE:ITEM:EFFICIENCY 1
Syntax
:MEASure:ITEM:EXTernalin
<NR1>
<NR1> = 0 - 7
Example
:MEASURE:ITEM:EXTERNALIN 7
As the default output items for
measurement by external signal input
unit, specify EXTA, EXTB, PM.
Function
Sets the default items (only efficiency
data) to be transferred in the response
message to the :MEASure? query in the
default mode.
The item is set as shown below by setting
bits, to specify a single numerical value.
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
EFF3 EFF2 EFF1
Error
Execution error / If the setting data is out
of the range.
Function
Queries the output item of the efficiency
measurement specified by the
:MEASure:ITEM:EFFiciency <NR1>
Function
Sets the default items (only data for the
external signal input unit) to be
transferred in the response message to the
:MEASure? query in the default mode.
The item is set as shown below by setting
bits, to specify a single numerical value.
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
PM EXTB EXTA
Error
Execution error / If the setting data is out
of the range.
:MEASure:ITEM:EFFiciency
Sets the output item of the efficiency measurement.
:MEASure:ITEM:EFFiciency?
Queries the output item of the efficiency measurement.
:MEASure:ITEM:EXTernalin
Sets the output item for the measurement value by using the 9603 External signal input unit.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......