184
────────────────────────────────────────────────────
:RS232c:HANDshake? (Queries for the RS-232C interface)
────────────────────────────────────────────────────
Syntax
:RS232c:HANDshake?
Response
syntax
Headers: ON
:RS232C:HANDSHAKE
<X/ HARD/
OFF>
Headers: OFF
<X/ HARD/ OFF>
Example
Transmission
Response
:RS232:HAND?
:RS232C:HANDSHAKE X
(headers on)
X
(headers off)
Syntax
:RS232c?
Response
syntax
Headers: ON
:RS232C:HANDSHAKE
<X/ HARD/
OFF>;
ANSWER
<ON/ OFF>
Headers: OFF
<X/ HARD/ OFF>;<ON/ OFF>
Example
Transmission
Response
:RS232?
:RS232C:HANDSHAKE OFF;ANSWER OFF
(headers on)
OFF;OFF
(headers off)
Syntax
:SCALe:
[1 - 6]
:CONTrol
<ON/OFF>,<ON/OFF>,<ON/OFF>
Example
:SCALE1:CONTROL ON,ON,OFF
Sets the PT and CT ratios to on, and
set the SC ratio to off in scaling
function for channel 1.
Function
The value of the communications
handshake is returned as character data
(X, HARD or OFF).
Errors
・
Query error/ If the response message is
longer than 2000 bytes.
・
Execution error/ If this query is executed
when using GP-IB interface
Notes
・
With this query, if any error occurs, no
response message is produced.
・
Even if a system error occurs, this query
is executed.
Function
Returns the current setting for the RS-
232C communications handshake as X,
HARD or OFF, the setting for execution
confirmation message as ON or OFF .
Errors
・
Query error/ If the response message is
longer than 2000 bytes.
・
Execution error/ If this query is executed
when using GP-IB interface
Notes
・
With this query, if any error occurs, no
response message is produced.
・
By using the TRANsmit:SEParator
command, the message unit separator can
be changed from the semicolon ";" to the
comma ",".
・
Even if a system error occurs, this query
is executed.
Function
Enables and disables PT, CT, SC ratios
scaling function for specified input unit.
Notes
In 1P3W mode or above, for the channel
specification, enter the number of the
lowest-numbered channel in the
combination of input units. For example,
when using input unit channels 1 to 3 in
3V3A mode, specify "1".
:RS232c:HANDshake?
(Queries for the RS-232C interface)
Queries the setting for the RS-232C communications handshake.
:RS232c?
(Queries for the RS-232C interface)
Queries the setting for the RS-232C communications handshake.
:SCALe [
channel no.
] :CONTrol
Enables and disables scaling function of PT, CT, SC ratios.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......