WARNING
In order to avoid electric shock or a short circuit, turn off the power
meter and the power flowing through the line being measured before
connecting or disconnecting a connector and an output connector.
CAUTION
・ To avoid damage to the unit, do not short the output terminal and do
not input voltage to the output terminal.
・ When not using the connector, to avoid damage to the main unit,
always fit the supplied cap.
・ The output terminals and control terminals are not insulated. Handle
these items carefully in order to avoid electric shock or a short circuit
accident.
NOTE
Chapter 10
D/A Output
10.1 Overview
This unit is provided with eight channels of D/A output as standard
equipment. The items displayed on the screen are output as DC voltages.
・
The output impedance is approximately 100
Ω
. When connecting to a
recorder, DMM, or similar, use a unit with a high input impedance (at least
1 M
Ω
).
・
For the specifications, see Chapter 20.
・
The outputs correspond to the values shown on the screen. In the hold state,
the displayed value continues to be output.
When both hold and interval settings are made, the output is updated each
interval time.
・
When the averaging function is activated, the averaged value is output.
・
Data from the harmonic analysis/flicker function is not output.
・
For the output circuit, see Section 9.2, "Internal Circuit for Analog, Monitor,
D/A Outputs."
・
On the STATUS or FDD screen, the output is not updated.
・
Data measured on the 9605 cannot be output.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......