235
────────────────────────────────────────────────────
14.3 Specifications (using with the 3193)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Voltage (
U
)
Current (
I
)
Active power (
P
)
Input type
Resistor voltage divider
+ isolation amplifier,
for isolated input
Isolated input, using DC-
CT method
――――
Measurement type
Analog processing:
・
True effective value
measurement
・
Mean value
rectification effective
value measurement
・
DC measurement
Analog processing:
・
True effective value
measurement
・
Mean value rectification
effective value
measurement
・
DC measurement
Analog processing:
Active power
measurement
Input resistance
2 M
Ω±
100 k
Ω
(
±
5%) 1
mΩ
or less
――――
Measurement range
6.0000/15.000/30.000
60.000/150.00/300.00
600.00 V /1.0000 kV
200.00/500.00 mA
1.0000/2.0000/5.0000
10.000/20.000/50.000 A
Depending on
combination of voltage
and current values.
Effective input range
5% to 110% (up to 100% for 1 kV range) or within the range accuracy
assurance range for the signal frequency specified elsewhere
Display range: 0.2% to 130% (AC/AC+DC, 200 mA range),
0.2% to 130% (others), 0.3% to 130% (1000 V range)
Maximum input voltage
Maximum input current
(55 Hz)
1000 Vrms
1500 Vpeak value
65 Arms
100 Apeak max.
――――
Crest factor
(Measurement range
×
6) / measurement value or
Maximum input peak value/ measurement value
――――
Accuracy
DC
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
(Note 1)
0.5 - 1 Hz
±
0.5% rdg.
±
0.5%f.s
±
0.5% rdg.
±
0.5%f.s.
±
0.5% rdg.
±
0.5%f.s.
23℃±5℃
(73°
F±9°
F)
80%RH max.
Power
factor=1
1 hour warm
-up
Input sine
wave
Voltage to
earth 0 V
Guaranteed
accuracy
period: six
months
After
degaussing
1 - 10 Hz
±
0.2% rdg.
±
0.2%f.s.
±
0.2% rdg.
±
0.2%f.s.
±
0.2% rdg.
±
0.2%f.s.
10 - 45 Hz
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
45 - 66 Hz
±
0.1% rdg.
±
0.1%f.s.
±
0.1% rdg.
±
0.1%f.s.
±
0.1% rdg.
±
0.1%f.s.
66 Hz
- 10 kHz
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
10 k -
50 kHz
±
0.3% rdg.
±
0.3%f.s
±
0.3% rdg.
±
0.3%f.s.
±
0.3% rdg.
±
0.3%f.s.
50 k
- 100 kHz
±
0.5% rdg.
±
0.5%f.s.
5 A or less
5 A or more
5 A or less
5 A or more
±
0.5% rdg.
±
0.5%f.s.
±
2.5%f.s.
±
0.5% rdg.
±
0.5%f.s.
±
5%f.s.
100 k -
300 kHz
±
0.5% rdg.
±
0.5%f.s.
±
0.5% rdg.
±
0.5%f.s.
±
5.0%f.s.
±
1.0% rdg.
±
1.5%f.s.
±
10%f.s.
300 k -
400 kHz
±
1.5% rdg.
±
0.5%f.s.
±
2.0% rdg.
±
0.5%f.s.
――
±
2.0% rdg.
±
1.5%f.s.
――
400 k -
500 kHz
±
2.0% rdg.
±
1.0%f.s.
±
2.0% rdg.
±
1.0%f.s.
±
2.0% rdg.
±
2.5%f.s.
500 k -
700 kHz
±
10.0%f.s.
±
10.0%f.s.
±
15.0%f.s.
700 k -
1 MHz
±
15.0%f.s.
±
15.0%f.s.
±
30%f.s.
14.3 Specifications (using with the 3193)
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......