275
────────────────────────────────────────────────────
20.2 Function Specifications
────────────────────────────────────────────────────
14. Other functions
Coupling mode switching
DC/ AC+DC/ AC
DC or AC+DC mode cannot be used when using the 9601 or
when using the 9602 with the AC current sensor
Rectifier type switching
RMS/ MEAN
In DC coupling mode, switching is not possible
Response switching
FAST/MID/SLOW
LPF switching
OFF/ 500 Hz/ 5 kHz/ 300 kHz (for the specifications, according to
each unit) For the 9601, OFF/500 Hz selection
Polarization detection
stabilization filter
OFF/ 200 Hz
It is effective when the calculation (type1) is selected.
Scaling
Displays (PT ratio, CT ratio, SC constant)
×
measurement value
Constant value: ".0001" to "10000"
Hold function
Pressing the HOLD key stops updating the all measurement value
display
Updates by pressing the key
Operation by time setting
Display update by external control
Peak hold function
In the peak hold state, the maximum value is updated from that
point
Time settings
Interval time (10 seconds to 100 hours) 10 second step
When using with the FDD or printer, depending on the number of
output times, minimum interval time is automatically changed.
Timer control (1 minute to 10000 hours) 1 minute step
Real time control 1 minute
Averaging
Time averaging (averaged by interval time, timer, real time control
which has been set)
Moving averaging (number of sampling: 8/16/32/64)
Exponential averaging (Attenuation constant 8/16/32/64)
Real time display
Displays year/month/day/hours (24-hour)/ minutes/seconds
accuracy:
±
25 ppm
±
1 second (25
o
C)
Battery back-up
Backup the settings and integration data
When the power failure occurs during averaging, integration, or
printing, restarts after restoring power failure.
Battery life 10 years or more (at 25
o
C reference value)
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......