183
────────────────────────────────────────────────────
:RS232c:ERRor? (Queries for the RS-232C interface)
────────────────────────────────────────────────────
Syntax
RS232c:ERRor?
Response
syntax
(Headers: ON)
:RS232C:ERROR
<0-7>
(Headers: OFF)
<0-7>
Example
Transmission
Response
:RS232:ERR?
:RS232C:ERROR 4
An overrun error has occurred.
Syntax
RS232c:HANDshake
<X/HARD/
OFF>
X: software handshake
HARD: hardware handshake
OFF: no handshake
Example
:RS232:HAND X
Sets to software handshake (X
parameter).
Function
・
Returns the RS-232C communications
error information as a numerical data
value in NR1 format (0 to 7) and then
clears.
・
Error information until reading by this
command from starting the RS232C
communication. Changing the
communication setting clears.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1 bit 0
Unused Unused Unused Unused Unused Overrun Framing Parity
・
Overrun error
・
Framing error
・
Parity error
Errors
・
Query error/ If the response message is
longer than 2000 bytes.
・
Execution error/ If this query is executed
when using GP-IB interface
Notes
・
With this query, if any error occurs, no
response message is produced.
・
Even if a system error occurs, this query
is executed.
・
This query clears the communications
error information.
Function
Select the type of handshake.
Errors
・
Execution error/ When using the GP-IB
interface
・
Command error/ If the setting data is not
character data other than X, OFF, HARD
Notes
Even if a system error occurs, this
command is executed.
:RS232c:ERRor?
(Queries for the RS-232C interface)
Queries whether or not the RS-232C communications error information are enabled.
:RS232c:HANDshake
(Command for the RS-232C interface)
Sets the RS-232C communications handshake
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......