9.2 Internal Circuit for Analog, Monitor, D/A Outputs
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Output type
Output rate
Analog Output
5Vf.s. (1kV range: 3.333Vf.s.)
Monitor Output
1Vf.s. (1kV range: 0.667Vf.s.)
D/A Output
5Vf.s.
CAUTION
To avoid damage to the unit, do not short the output terminal and do
not input voltage to the output terminal.
Output terminal
Analog GND
NOTE
9.2 Internal Circuit for Analog, Monitor, D/A Outputs
The output rates are shown below.
For the details, please see the specifications of the output unit.
・
The output impedance is approximately 100
Ω
. When connecting to a
recorder, DMM, or similar, use a unit with a high input impedance (at least
1 M
Ω
).
・
See the Specifications of each input units for the analog output and monitor
output.
・
For the specification of the D/A output, see chapter 10, "D/A Output."