APPENDIX 1
────────────────────────────────────────────────────
────────────────────────────────────────────────────
Voltage
Load
②
①
Diagram for the example
using the option 9605
Appendix
3P3W wiring
①
Voltage is measured between lines and current is measured with phase, so
U2 is phased 60
°
from U1. When the load is purely resistive, I1 lags U1 by
30
°
, and I2 leads U2 by 30
°
.
②
With an inductive load such as a motor, when the effective power of ch1 is
negative, the phase of I1 lags U1 by more than 90
°
.