A
NOTE
Range
500.00 mHz
to 50.000 Hz
20.00 Hz to
500.00 Hz
200.00 Hz to
5.0000 kHz
2.0000 kHz to
50.000 kHz
20.000 kHz to
2.0000 MHz
HPF
0.5 Hz
100 Hz
10 kHz
LPF
360 Hz
50 kHz
1.2 MHz
51.tif
Frequency
measurement data
Chapter 5
Frequency Measurement
This unit has internal circuits for three frequency measurement channels (fa,
fb, fc), and can thus measure a number of systems simultaneously.
The frequency ranges can be combined with high-pass filters (HPF) and low-
pass filters (LPF)
・
Depending on the frequency range and the frequency and distortion of the
waveform, stable measurement may not be possible.
In that case, set the range manually.
・
The frequency effective measurement range is within the range of frequency
characteristics of combination input unit.
・
When using the 9603
EXTERNAL SIGNAL INPUT UNIT
you select pulse
measurement, this is unconditionally assigned to the 9603 unit. For details
see Chapter 17, "9603
EXTERNAL SIGNAL INPUT UNIT
."
(9603 has no HPF and LPF)
The frequency measurement data is displayed on the lower right of the
"DETAILS" page on each channel screen
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......