159
────────────────────────────────────────────────────
:DISPlay:MAGnify [channel no.]
────────────────────────────────────────────────────
Syntax
:DISPlay:MAGnify
[1 - 6] <A,..>
Example
When measuring in 3V3A mode with
using the channels 1 to 3 of the input
unit.
:DISPLAY:MAGNIFY1 U123,
I123,P123,PF123
Sets the display item to U123, I123,
P123, PF123 (power factor
λ
123) on
the enlarged display for channels 1 to
3.
Syntax
:DISPlay:MAGnify
[1 - 6]
?
Response
syntax
:DISPLAY:MAGNIFY
[1 - 6] <A>
Example
Transmission
Response
:DISPLAY:MAGNIFY1?
:DISPLAY:MAGNIFY1
U123,I123,P123,PF123
Function
Sets items to be displayed on the
enlarged screen for the specified channel
If the data is not specified, the enlarged
screen for the specified channel is
displayed.
Errors
Execution error / If the current integration
value or power integration value is
specified, or if the setting data is other
than character data.
Note
・
For the detail screen for is 1P3W or
above, for the channel specification enter
the number of the lowest-numbered
channel in the combination. For example,
when using input unit channels 1 to 3 in
3P4W mode, specify "1".
・
The number of character data
<
A
>
which
can be specified is up to 4 items and the
items are displayed in specified order.
The items varies depending on the
connection mode. See :DISPlay:DETail
・
If the setting exceeds maximum number
of set item, the data exceeded is ignored.
・
When "Pk" (waveform peak value) is
specified, the peak value which is set by
the :WAVEPeak[ch] command (switching
voltage/current) is displayed.
・
"PF" (power factor
λ
) and "DEG"
(phase angle
φ
) cannot be displayed
simultaneously. If set both, the latter is
displayed.
・
For the frequency measurement values
(FA, FB, FC), if a channel is specified
which is not related to the specified
screen display channel, the specification
cannot be made.
Function
Queries items to be displayed on the
enlarged screen for the specified channel.
Note
For the detail screen for is 1P3W or
above, for the channel specification enter
the number of the lowest-numbered
channel in the combination. For example,
when using input unit channels 1 to 3 in
3P4W mode, specify "1".
:DISPlay:MAGnify [
channel no.
]
Sets items to be displayed on the enlarged screen for the specified channel
:DISPlay:MAGnify [
channel no.
]?
Queries items to be displayed on the enlarged screen for the specified channel.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......