252
────────────────────────────────────────────────────
16.4 Internal Block Diagram
────────────────────────────────────────────────────
U
±
±
Attenuation
circuit
Buffer
Range
Range
Isolation
amplifier
HPF
LPF
RMS/DC
MEAN/DC
A/D
3193
3193
-10
Switch control
RMS/DC
MEAN/DC
Multiplication
/DC
HPF
LPF
Auto-zero
Peak hold
Waveform
shaping
Polarity
decision
Over
detect
f measurement
U
/
I
D/A
LEAD/LAG
OVER
I
DMAG (0 Adj)
Sensor
OVER
U
16.4 Internal Block Diagram
The voltage value is converted by attenuator and range circuits to a voltage
signal waveform proportional to the measured voltage, then isolated by an
isolating amplifier.
The sensor input is converted in a range circuit to a voltage signal waveform
proportional to the measurement current.
For details on RMS value, MEAN value, active power, waveform peak value
measurement circuit, peakover detect circuit, see Chapter 14.