272
────────────────────────────────────────────────────
20.2 Function Specifications
────────────────────────────────────────────────────
③ Motor power (
P
m measurement)
Measurement type
Digital computation from the measured voltage or pulse signal.
In the case that the 9603 channel A unit is torque and channel B
unit is rotation count/rate.
Accuracy
±
1dgt. with respect to the value computed from the measurement
values
Display range
0.1 to 130% of setting voltage range when both channel A and B
is set to DC voltage measurement
0 to 100% of using frequency measurement range when channel B
is set to pulse measurement.
Unit indication
W
Calculation
The calculation is carried out as follows, depending on the torque
units set for channel A.
Unit of
chA
Motor power calculations (W)
N・m
2×π×(display value of chB)
(Display value of chA)× ─────────────
60
mN・m
2×π×(display value of chB)
(Display value of chA)× ─────────────
60×100
kN・m
2×π×(display value of chB)×1000
(Display value of chA)× ────────────────
60
kgf・m
2×π×(display value of chB)×9.80665
(Display value of chA)× ────────────────
60
kgf・cm
2×π×(display value of chB)×9.80665
(Display value of chA)× ────────────────
60×100
7. Frequency measurement (
fa, fb, fc)
(when using the 9600, 9601, 9602)
Measurement type
Reciprocal calculation from frequency of input waveform
Number of measurement
channel
3 channels
Display range
AUTO/ 50 Hz/ 500 Hz/ 5 kHz/ 50 kHz/ 2 MHz
Resolution
99999 counts
Measurement range
500.00 mHz to 50.000 Hz/ 20.000 Hz to 500.00 Hz/
200.00 Hz to 5.0000 kHz/ 2.0000 kHz to 50.000 kHz/
20.000 kHz to 2.0000 MHz
Measurement accuracy
±
0.1% rdg.
±
1dgt.
With a sine wave input from 10% to 130% of the
U
/
I
ranges,
when the frequency characteristics of the input unit forming the
source are within the specified range
Function
Switching
U
/
I
source
(Simultaneous
U
/
I
input from a single input unit is not possible.)
Filter switching linked to range (high-pass or low-pass filter)
Pulse measurement when 9603 is used
When the 9303 is set to pulse measurement, fc is forced to pulse
measurement of channel B.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......