273
────────────────────────────────────────────────────
20.2 Function Specifications
────────────────────────────────────────────────────
16
17
18
19
20
6
7
8
9
10
11
12
13
14
A
8. Integration measurement (current integration
Ih, power integration WP)
Measurement type
Digital calculation from the measured value of
I
,
P
Accuracy
±
1dgt. with respect to the value computed from the measurement
value for each channel (
I
,
P
)
Number of measurement
64 times/s
Measurement item
Integration of current and active power for all channels is possible
The following items depends on settings
1P2W, DC mode:
+Ih
,
-Ih
,
I h
,
+WP
,
-WP
,
WP
1P2W, excluding above:
I h
,
+WP
,
-WP
,
WP
1P3W or more: Ih for each channels, +
WP
, -
WP
,
WP
of SUM
value for active power
Measurement range
0 to
±
9999999TAh/TWh (when integration time is within 10,000
hours)
Effective input range
Same as effective input range of each input units
Integration time accuracy
±
25ppm
±
1dgt.
(
0
o
C to 40
o
C
)
Function
Separate integration for each polarity (positive, negative, and total)
Integration for various control times
Automatic output to FDD/printer
Display of integration elapsed time
Cumulative integration by repeated start/stop operations
Start/stop/reset by external control
(Simultaneous control of selected channels; independent control
not possible)
Back-up of the integration value and integration elapsed time
during power failure
Restart of integration after restoring power failure
9. Load factor measurement function (
LF)
Calculation
Digital calculation from the positive integration value of active
power (only when the interval time is set)
Accuracy
Same as integration measurement
Number of calculation
Same as integration measurement
Measurement range
0.00 to 100.00%
Calculation
W
av
Load factor (
LF
) =
───
×
100%
W
max
W
av : Total averaging
W
max : Maximum value of averaging for interval
Total integration value (+)
W
av =
───────────────
Timer time or real-time control time
Maximum value of integration value for interval
W
max =
──────────────────
Interval time
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......