140
────────────────────────────────────────────────────
*TRG
────────────────────────────────────────────────────
Syntax
*
TRG
Syntax
*
TST?
Response
syntax
(Header ON/OFF)
<NR1>
<NR1> = 0 to 31
Example
Transmission
Response
*TST?
20
(Input unit error + printer error)
Syntax
*
WAI
Example
:VOLT1:RANG 150;:RESP
SLOW;*WAI;:MEAS? U1,I1
After operation for voltage range and
response, the :MEAS? command is
executed.
Function
・
Same operation as the :HOLD command.
・
If the system is currently in the hold
state, performs sampling once.
Function
Causes the 3193 to perform the self test,
and returns the result thereof as a
numerical data value in NR1 format (0 to
31).
128
64
32
16
8
4
2
1
bit 7 bit 6 bit 5 bit 4 bit 3
bit 2
bit 1 bit 0
Printer
error
RTC
error
Input unit
error
RAM
error
ROM
error
・
Input unit error:
no sensor of the clamp unit, no unit for
channel 1
・
RTC error:
internal clock abnormal
・
Printer error:
no printer paper, head-up, temperature
out of range
Note
・
No header is affixed to the response
message.
Function
When a sequence of commands is written
in a single line, the process waits until all
commands before the *WAI command
have completed before continuing with
the following commands.
*
TRG
Request for sampling
*
TST?
Requests execution of, and queries the result of, the self test.
*
WAI
Waits until sampling is fully completed.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......