243
────────────────────────────────────────────────────
15.3 Specifications (using with the HIOKI 3193)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Voltage (
U
)
Current (
I
)
Active power (
P
)
Input type
Resistor voltage divider
+ isolation amplifier,
for isolated input
Isolated input, using CT
method
――――
Measurement type
Analog processing:
・
True effective value
measurement
・
Mean value
rectification effective
value measurement
Analog processing:
・
True effective value
measurement
・
Mean value rectification
effective value
measurement
Analog processing:
Active power
measurement
Input resistance
2 M
Ω±
100 k
Ω
(
±
5
%
) 1
mΩ
or less
――――
Measurement range
6.0000/15.000/30.000
60.000/150.00/300.00
600.00 V/1.0000 kV
200.00/500.00 mA
1.0000/2.0000/5.0000
10.000/20.000/50.000 A
Depending on
combination of voltage
and current values.
Effective input range
5% to 110% (up to 100% for 1 kV range) or within the range accuracy
assurance range for the signal frequency specified elsewhere
(
display range 0.1% to 130%(200 mA range), 0.2% to 130%(Others),
0.3% to 130%(1000V range)
)
Maximum input voltage
Maximum input current
(55 Hz)
1000 Vrms
1500 Vpeak value
65 Arms
100 Apeak max.
――――
Crest factor
(Measurement range
×
6)
/measurement value or
Maximum input
voltage peak value/
measurement value
(Measurement range
×
6)
/measurement value or
Maximum input current
peak value/ measurement
value
――――
Accuracy
(Note 1)
23℃±5℃
(73°
F±9°
F)
80%RH max.
Power
factor=1
1 hour warm
-up
Input sine
wave
Voltage to
earth 0 V
Guaranteed
accuracy
period: six
months
5-10 Hz
±
2.5%f.s
±
2.5%f.s
±
2.5%f.s
10 - 20 Hz
±
1.0%f.s
±
1.0%f.s
±
1.0%f.s
20 - 45 Hz
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
45 - 66 Hz
±
0.1% rdg.
±
0.1%f.s.
±
0.1% rdg.
±
0.1%f.s.
±
0.1% rdg.
±
0.1%f.s.
66 Hz
- 5 kHz
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
±
0.1% rdg.
±
0.2%f.s.
5 k -
10 kHz
±
0.2% rdg.
±
0.4%f.s
±
0.2% rdg.
±
0.4%f.s.
±
0.2% rdg.
±
0.4%f.s.
10 k -
20 kHz
±
1.0%f.s.
±
1.0%f.s.
±
1.0%f.s.
20 k -
50 kHz
±
2.5%f.s.
±
2.5%f.s.
±
2.5%f.s.
50 k -
100 kHz
±
10.0%f.s.
±
10.0%f.s.
±
10.0%f.s.
Note 1: There are limited ranges for the levels at which the accuracy is specified, depending on the input
signal frequency.
Note 2: The accuracy specified range varies depending on switching response, coupling mode, and LPF.
See each column on the next page.
15.3 Specifications (using with the HIOKI 3193)
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......