79
────────────────────────────────────────────────────
8.1 Overview
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
NOTE
Chapter 8
Efficiency Measurement
8.1 Overview
This unit can calculate the efficiency from the measured values (active
power, motor power). For example, the input/output efficiency of an inverter,
input/output efficiency of a motor, and overall efficiency can be calculated
simultaneously with a single unit.
・
Measuring the motor power (
P
m) requires the optional 9603 EXTERNAL
SIGNAL INPUT UNIT.
・
If the load is subject to violent fluctuations or transients, stable measurement
may not be possible.
・
The coupling mode on the channels for which the efficiency is to be
measured should be the same. (DC/AC + DC/AC)
・
If the calculated efficiency exceeds 100%, it is treated as 100%.