4.6 Setting the Low-pass Filter (LPF)
The input units of the 3193 are provided with a low-pass filter function for
restricting the frequency characteristics.
By using an appropriate filter selection it is possible to eliminate harmonics.
Depending on using the input units, the low-pass filter (LPF) may not be
selected. The frequency range of accuracy assured varies. For details, see
the specifications of the input units.
Switching using the panel keys
1. Switch to display the channel for which you wish
to change the setting.
2. Press the
SHIFT
key, then use the
CURSOR
key to
change.
Pressing the
CURSOR
key switches low-pass
filter.
Switching on the STATUS screen
1. Press the
STATUS
key, then use the
PAGE
(
)
keys to display the "
UNIT
" page.
2. Using the
CURSOR
keys, move the cursor to the
channel to be changed of "
LPF
" item.
3. Select desired low-pass filter from
F1
(OFF)
,
F2
(
500 Hz
),
F3
(
5 kHz
),
F4
(
300 kHz
).
4. Press the
MEAS
key to return to the measurement
screen.
47-3.tif
47-2.tif
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......