178
────────────────────────────────────────────────────
:MEASure?
────────────────────────────────────────────────────
Syntax
①
Default mode
:MEASure?
②
Data specification mode
:MEASure?
<A (up to 70 items)>
<A>= U1 - U6, U12, U34, U56, U45, U123, U456
I1 - I6, I12, I34, I56, I45, I123, I456
P1 - P6, P12, P34, P56, P45, P123, P456
S1 - S6, S12, S34, S56, S45, S123, S456
Q1 - Q6, Q12, Q34, Q56, Q45, Q123, Q456
PF1 - PF6, PF12, PF34, PF56, PF45, PF123, PF456
DEG1 - DEG6, DEG12, DEG34, DEG56, DEG45,
DEG123, DEG456, Pk1 - Pk6, FA, FB, FC
EFF1, EFF2, EFF3, EXTA, EXTB, PM
PIH1 - PIH6, MIH1 - MIH6, IH1 - IH6
PWP1 - PWP6, PWP12, PWP34, PWP56, PWP45,
PWP123, PWP456, MWP1 - MWP6, MWP12,
MWP34, MWP56, MWP45, MWP123, MWP456
WP1 - WP6, WP12, WP34, WP56, WP45, WP123,
WP456, TIME, LF1 - LF6, LF12, LF34, LF56, LF45,
LF123, LF456
Response
syntax
Headers on
(
header1) <A1>
;
(header2)
<A2>
;...;
(header70) <A70>
?
Headers off
<A1>
,
<A2>
,..,
<A>
?
Example
Transmission
Response
:MEASURE? U1,I1,P1
U1 10.230E+00;I1 1.000E+00;
P1 1.340E+03
(headers on)
10.230E+00;1.000E+00;1.340E+03
(
headers off)
Function
①
Default mode
If no parameters are specified in the data
section, then this mode is used. Default
item data specified by the
:MEASure:ITEM command is created. In
this case the data order is fixed.
②
Data (parameter) specification mode
If one or more parameters are specified in
the data section, then this mode is used.
Measurement item data specified by <A>
is created. The order of arranging the
data (parameters) is freely selectable, and
data is created in the specified order.
Header
Data portion
Excluding
Integration
value
Numerical data in NR3 format
±□□□□□□E±□□
Mantissa : 6 digits with a decimal point
Exponent : 2 digits
Integration
value
Numerical data in NR3 format
±□□□□□□□E±□□
Mantissa : 6 digits with a decimal point
Exponent : 2 digits
Time
Numerical data in NR1 format
□□□□□,□□,□□
(hours, minutes, seconds)
Error
Display blank
+6666.6E+99
Scaling error
+7777.7E+99
Input over
+9999.9E+99
Note
・
Up to 70 items can be responsed.
・
To change the NR3 numerical data
format, see the :TRANsmit:COLumn
command.
:MEASure?
Queries the specified data.
Summary of Contents for Power HiTester 3193
Page 2: ......
Page 50: ...32 3 9 Operations During Power Failure ...
Page 76: ...58 4 13 Degaussing ...
Page 80: ...62 5 2 Setting the Frequency Range fa ...
Page 108: ...90 9 3 Internal Circuit for the External Control and Timing ...
Page 112: ...94 10 3 Output Rate ...
Page 250: ...232 13 9 Error and Overflow Displays ...
Page 278: ...260 17 5 Internal Block Diagram ...
Page 284: ...266 19 2 Installation Procedures For JIS standard For EIA standard External Dimensions ...
Page 300: ...282 20 4 Internal Block Diagram of the 3193 ...
Page 306: ...INDEX 4 Index ...
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......