Direct Control and Replenishment
338
CEQ™ 8000 Genetic Analysis System
8. Grasp the Array Fitting tab (Figure 203) and then:
a. Pull the fitting approximately one inch out of the manifold.
b. Touch the tip of the fitting to the bottom of the Optics Base Plate.
c. Hold and wait five seconds for the gel strand to dry.
d. Pull the fitting away from the instrument.
e. Wipe gel strands off of the instrument using a damp tissue.
Note the number of runs and days on the instrument for this capillary
array for reference. If this capillary array is to be re-installed, this
information is necessary. A label has been provided on the back of the
array packaging container so that this information may be recorded.