Routine Maintenance
User’s Guide
353
d. Pull the fitting away from the instrument.
e. Use a tissue to wipe gel strands off of the instrument.
Figure 217:Removing/Replacing the Array Fitting
8. With the array fitting in hand, blow dust and debris off of the windows with
compressed gas (Texwipe Microduster III, P/N: TX2511).
9. Using a water-moistened swab (Texwipe Swab, P/N: TX754B), gently wipe the
Detection Window by stroking in one direction
only
as shown in Figure 218.
900489e.AI
900507e.AI
900489e .AI
Guide Pins
Eject Lever
Array Fitting
900507e .AI
gel strand
Optics Base Plate
Array Fitting
(bottom)