Direct Control and Replenishment
340
CEQ™ 8000 Genetic Analysis System
Always grip the capillary array fitting near the end during removal
or installation of the tip cap to prevent flexing and possible
breakage of the capillary array.
8.
While holding the Array Fitting tab (of the new capillary array), align the Array
Fitting (Figure 203) with the manifold opening and guide pins. Push the fitting into
the manifold until it is completely seated against the bases of the Guide Pins. Make
sure that the fitting tab is placed downward when inserting the capillary array into
its slot.
9.
While holding the Electrode Block tab, align the Electrode Block (Figure 204)
with the guide block pins and gently push it in until resistance is met. (The
resistance is from the spring-loaded contacts.)
10. When installing the capillary array, carefully route the capillaries through the hole
in the Plenum Assembly (Figure 205) and straight across into the guide block pins
(Figure 206).
CAUTION
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...