When reinstalling the Plenum Assembly, gently gather the capillaries
between your thumb and forefinger to make sure they pass through the
hole in the Plenum Assembly without any constriction.
11. Replace the Plenum Assembly and tighten the two captive screws (Figure 202).
12. Replace the Manifold Access Cover and tighten the captive screw (Figure 201).
13. Lower the Capillary Temperature Control Cover and secure the two rubber latches.
14. Lower the Capillary Access Cover and Sample Access Cover to their locking
positions
CAUTION:
USE THIS FRONT PLENUM
ONLY WITH A 33 CM
CAPILLARY ARRAY.
PRINED IN U.S.A.
608542-AA
900697L.AI
900694L.AI
HV
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...