System Overview
16
CEQ™ 8000 Genetic Analysis System
EMI Effect on CEQ™ (2000 Series / 8000 Series) Performance and Recommended
Mitigations
Under the test conditions specified by the European normative electromagnetic
compatibility standard EN 61326-1, the CEQ instrument may exhibit temporary
degradations in performance in accordance with the table below.
Because the environmental circumstances contributing to the problem can vary,
several different mitigation techniques have been provided that should help eliminate
or reduce the interference.
Test Condition
Effect on Performance
Mitigation
Radio Frequency
Field
Interference (RFI)
System exposed
to electromagnetic
field strengths of
greater than or
equal to 3 V / m at
multiple frequency
bands.
May cause temporary degradation
of base calling accuracy. In some
cases, significant noise will appear
in the measured results. Will
return to normal performance
once exposure is removed.
• Re-run sample.
• Move location of the product
by several meters.
• Change the orientation of the
equipment by 90 degrees.
• Avoid using transmitters or
cellular phones within 1 meter
of the equipment.
• Review the temperatures in
the run log for any anomalies
and re-run as necessary.
May cause temporary loss of
capillary temperature control
resulting in lower than expected
capillary temperature. Affect will
be documented in the run log. Will
return to normal performance
once exposure is removed.
May cause temporary loss of
capillary temperature control
resulting in run not being initiated
because temperature cannot
reach desired level.
• Apply mitigation techniques
above.
• Disabling the temperature
limit feature in the Method Edit
screen (Capillary
Temperature box) will allow
the run to be performed
without desired temperature
being reached.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...