Routine Maintenance
352
CEQ™ 8000 Genetic Analysis System
Cleaning the Capillary Array
The capillary array windows must kept free of any contaminants.
Otherwise, high backgrounds and/or drifting baselines will occur.
All background levels should be below 6000 counts.
Water used during this procedure must be fresh, distilled, deionized
water (18 Mohm/cm water).
1. Select
Replenish | Release Capillary Array
from the Run Module Main
window. Wait for the Remove Capillary Array dialog to appear.
2. Open the Sample Access Cover (Figure 215) and lift to the vertical locking
position.
3. Open the Capillary Access Cover and lift to the vertical locking position.
4. Unlatch the two rubber latches holding the Capillary Temperature Control Cover
and lift to the vertical locking position.
5. Loosen the Manifold Access Cover captive screw (Figure 216), remove the cover
and set it aside.
Figure 216:Manifold Access Cover
6. Lift the Eject Lever (Figure 217) to release the Array Fitting.
7. Grasp the Array Fitting tab (Figure 217) and then:
a. Pull the fitting approximately one inch out of the manifold.
b. Touch the tip of the fitting to the bottom of the Optics Base Plate.
c. Hold and wait five seconds for the gel strand to dry.
CAUTION
Manifold
Access
Cover
(removed)
900485e.AI
900487e.AI
900488e.AI
HV
LASER
PWR
900485e .AI
900487e .AI
HV
Capillary Temperature
Control Cover
(open)
Manifold
Access
Cover
captive screw
900488e .AI
Plenum
Assembly
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...