Biological Waste Disposal
User’s Guide
359
Bulk Disposal
When performing this procedure, use an exhaust ventilation unit
that meets TLV requirements.
1. Using a 1L side arm flask as a trap, connect the trap to a vacuum.
2. Attached a pipette to the trap using chemical resistant tubing.
3. Aspirate the formamide from each well and dispose of it in a hazardous
liquid organic waste container.
4. After complete removal of all formamide from the plates, allow the
empty plate to sit in the ventilation unit for 24 hours.
5. Dispose of the plate in a solid waste container.
Dispose of formamide in accordance with all applicable federal,
state and local environmental regulations concerning hazardous
liquid waste.
Disposal of Buffer/Gel Mixture from the Buffer Plate
Multi-Channel Pipettor
When performing this procedure, use an exhaust ventilation unit
that meets TLV requirements.
1. Aspirate 400 µL of buffer/gel mixture from the 96-well plate.
2. Dispose the buffer/gel mixture into a hazardous liquid organic waste
container.
3. Rinse the plate wells with water and dispose of the rinse in the liquid
waste container.
4. After all wells are clear, allow the empty plate to sit in the ventilation
unit for 24 hours.
5. Dispose of the plate in a solid waste container.
Dispose of buffer/gel mixture in accordance with all applicable
federal, state and local environmental regulations concerning
hazardous liquid waste.
WARNING
WARNING
WARNING
WARNING
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...