User’s Guide
233
Fragment Analysis Parameters - Analysis Method tab
This tab allows you to specify the size standard and model to be applied to the
calibration curve. The model selected should be the best fit for the standard, as it is
used to generate the calibration curve, which estimates sizes for the unknown
fragments.
Figure 136:Fragment Analysis Parameters window - Analysis Method tab
Size standards are sets of DNA fragments of known sizes that are run with a sample
and are used to create the calibration curve. This calibration curve is then used to relate
the known fragment size to either migration time or mobility in order to estimate the
size of the unknown fragment(s).
When creating a new size standard, there are two requirements that help the system to
find the standards and make correct associations between the fragment peaks and the
fragment list. Refer to Online Help for information on creating size standards.
1. Fragments in the larger half of the size range should be spaced uniformly.
•
For example, the peaks in the range of 200 to 400 nucleotides of the
SizeStandard-400 are 20 nucleotides apart.
2. Fragments in the smaller half of the size range should include a set of at least three
fragments that are spaced approximately half the distance of those that are in the
larger fragment size range.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...