Direct Control Procedures
User’s Guide
347
Figure 211:Install Gel Cartridge Dialog
5.
Dispose of the used tissue and spent gel cartridge in accordance with the
procedure,
“Disposal of the Gel Cartridge’
Removing the Manifold Plug
The Manifold Plug is installed in the CEQ instrument during shipping and when the
instrument is not is use to prevent drying of gel.
This procedure assumes that the Manifold Plug is being removed in
preparation for Capillary Array installation.
1. Select
Replenish | Release Capillary Array
from the Run Module main
menu (wait for the
Remove Manifold Plug
dialog box to appear).
2. Open the Sample Access Cover (Figure 190) and lift to the vertical locking
position.
3. Open the Capillary Access Cover and lift to the vertical locking position.
4. Unlatch the two rubber latches holding the Capillary Temperature Control Cover
and lift to the vertical locking position (Figure 212).
5. Loosen the Manifold Access Cover captive screw, remove the cover and set it
aside.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...