Direct Control Procedures
User’s Guide
343
Removing and Replacing a Gel Cartridge/Gel Pump Plug
Care must be exercised when installing/removing a gel car-
tridge due to the viscosity of the gel mixture. Do not allow
gel to remain on the instrument.
This procedure assumes that an expended gel cartridge is being
replaced with a fresh gel cartridge or that a used gel cartridge is
being removed (for storage purposes) and being replaced with the gel
pump plug.
1. Select
Replenish | Release Gel Cartridge
from the Run menu. When the
system is ready to release the gel cartridge, the
Release Gel Cartridge
dialog
box will be displayed. Wait until the lead screw is completely disengaged.
2. Open the Gel Pump/Gel Cartridge Access Cover by gently pushing in on the top of
the cover. (The cover is spring-loaded and will pop open.)
3. Pull on the Cartridge Locking Lever. (The barrel will swing outwards to
approximately a 90° angle from its locked position.)
CAUTION
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...