Fragment Analysis Parameters - Advanced Tab
The Advanced tab is used to set additional fragment analysis parameters. From this
window you can assign dye mobility correction values to compensate for the varying
effects of different dyes on fragment mobility, select a standard mobility reference and
indicate whether dye spectra used should be estimated from sample data or from saved
system spectra.
Figure 138:Fragment Analysis Parameters window - Advanced tab
Mobility
1. Select the desired
Dye Mobility Calibration
from the drop-down list box.
•
The Dye Mobility Calibration is used to select a specific calibration from
the options available in the list box and to adjust migration times or
mobility to compensate for the varying effects of different dyes on
fragment mobility.
•
Select
PA ver.1
if analyzing samples made from primers labeled with the
phosphoramidite dyes D2-PA, D3-PA, D4-PA.
•
AE ver.2
provides an improved calibration for primers labeled with the
active ester dyes (D1, D2, D3 or D4) and should be selected to obtain the
most accurate size estimates for these primers.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...