Running a Sample Plate
User’s Guide
143
You will be notified and the system will perform a gel volume re-estimation if you
attempt to run a plate when the total cartridge gel volume is less than 1.5 mL
more
than the minimum amount needed to process the plate.
If the system determines that there might not be enough gel to run the plate, it will
perform a 2-step re-estimation process automatically.
In the first step, the system will “home” the gel pump and the message in Figure 83
will appear.
Figure 83: Gel Volume Re-estimation - Step 1 Dialog
In the second step, the Gel Pump plunger will be engaged and the system will
accurately determine the gel volume.
The dialog box in Figure 84 will be displayed.
Figure 84: Gel Volume Re-estimation - Step 2 Dialog
If the gel volume is
greater
that 1.5 mL more than the minimum amount needed to
process the plate, the plate will run and the system will not re-estimate the gel volume.
Gel Re-Estimation Procedure
1. Launch the Run module to run a sample plate.
2. Select
Run | Start Sample Plate
to process the sample plate.
You can view the gel cartridge volume by clicking the
Life
tab on the Status
Monitor.
3. If the system determines that the gel cartridge volume is less than the volume
needed to safely process the plate, you will be presented with the two
re-estimation dialog boxes before the system starts running the sample plate.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...