Run Procedures
144
CEQ™ 8000 Genetic Analysis System
The system will measure the gel volume accurately and will determine whether
or not to proceed with the run.
4. If the actual volume (re-estimated by the system) is
greater
than 1.5 mL more than
the total volume needed to run the plate, the system will automatically start the
plate.
5. If the actual volume (re-estimated by the system) is
less
than 1.0 mL more than the
total volume to run the plate, you will be presented with the dialog box in
Figure 85.
Figure 85: Gel Usage Confirmation Dialog
6. Click
Accept
to proceed with the run. If the run is selected to continue, specify the
sample sets to run. Select
Replace Gel Cartridge
to replace the gel cartridge
before the run.
Gel Volume Notification
A dialog box will warn you if you attempt any operation requiring gel when there is
not enough gel in the cartridge to perform the operation.
You will be presented with the warning message in Figure 86 if you attempt to
perform a capillary gel fill when the cartridge does not have enough gel.
Figure 86: Gel Warning Dialog
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...