Direct Control and Replenishment
348
CEQ™ 8000 Genetic Analysis System
Figure 212:Manifold Access Cover
6. Lift the Eject Lever to release the Manifold Plug.
7. Grasp the Manifold Plug tab (Figure 213) and then:
a. Pull the plug approximately one inch out of the manifold.
b. Touch the tip of the plug to the bottom of the Optics Base Plate.
c. Hold and wait five seconds for the gel strand to dry.
d. Pull the plug out and set it aside for future use.
e. Wipe gel strands off of the instrument using a damp tissue.
Manifold
Access
Cover
(removed)
900485e.AI
900487e.AI
900488e.AI
HV
LASER
PWR
900485e .AI
900487e .AI
HV
Capillary Temperature
Control Cover
(open)
Manifold
Access
Cover
captive screw
900488e .AI
Plenum
Assembly
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...