User’s Guide
293
Iterations of similar data can be achieved using Loop delimiters or Tables. While the
data object names may appear squashed in the report template, if the data fits in the
space provided, the tables should look normal in the report.
Graph objects and Graph Placeholders must remain together for proper graph display.
Changing the graph display is discussed below.
Editing Graph Displays
Five features of the graph displays can be edited: 1) X scale type. 2) X-axis scale, 3)
Y-axis scale, 4) Pane Count or Pane Width and 5) Dyes displayed.
To Edit the graph:
1. Right click on the graph placeholder.
2. Select Format Object.
3. Go to the “Web” tab and create the following text:
<DisplayOptions
(Your commands will go here)
/>
Editing X scale type
The possible values for XScaleType are “Default”, “BySize”, and “ByTime”.
XscaleType only applies to Electropherogram view. “Default” allows the graph type to
determine whether to display by size or by time. If the XScaleType attribute is not
specified, it defaults to “Default”.
In the DisplayOptions section of the Alternative text: box, enter by size or time as
follows:
<DisplayOptions
XScaleType="BySize"
/>
OR
<DisplayOptions
XScaleType="ByTime"
/>
Editing the X-axis
X axis values specify either the size or the time, depending on the XScaleType. If
either or both are not specified, they take on the minimum and maximum values from
Unzoom all graph.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...