Quality Values
User’s Guide
165
•
the average Call Score/Quality Value in a window of 25 bases falls below 0.75 -
0.8 on a linear scale and 6 - 7 on a log scale,
and
•
there are at least 5 bases within that 25 base window whose Call Score/Quality
Value are less than 0.75 on a linear scale and 6 on a log scale.
Performing Quality-based Trimming
To setup Quality-based trimming, open the
Sequence Analysis
module and follow the
steps below.
1. Open a result and select the
Analysis | Working Analysis Parameters
menu
item.
2. Click the
Edit
button to open the
Sequence Analysis Parameters Editor
.
3. Select the
Quality-based Trimming
tab.
4. Select the
Ends To Be Trimmed
and the
Trimming Stringency
.
5. Click
Save As
and save the new parameters.
6. Click
OK
.
At this point, the result can be trimmed.
1. Select the
Analysis | Trim Based on Quality
menu item.
2. The Analysis Parameters, created in the steps above, should be listed. If not, click
Use Stored Parameters
and select the appropriate Analysis Parameter.
3. Click
OK
.
The resulting Base Sequence and Trace view displays the base letters, if poor quality,
as strikethrough after a Quality-based trim has been performed as shown in
Figure 100.
When performing Quality-based and Sequence-based trimming,
Quality-based trimming is always performed first.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...