Direct Control Procedures
User’s Guide
345
Figure 209:Removing/Replacing the Gel Cartridge - Continued
6. Remove any air pockets from the gel cartridge by grasping the cartridge wings
between your first two fingers, and depressing the plunger with your thumb until a
small amount of gel is pushed out of the cartridge tip.
If there are bubbles present in the gel, do not install the cartridge.
Repeat step 6 above before installing this cartridge.
7. Insert the new cartridge or the gel pump plug into the barrel and lock it into
position by aligning the cartridge wings with the cartridge holder and pushing in.
8. Push the cartridge locking lever towards the back of the instrument (approximately
a 90° angle) into its locked position.
9. Close the Gel Pump/Gel Cartridge Access Cover.
10. In the
Remove Gel Cartridge
•
Click
Install Cartridge
to indicate that a gel cartridge was installed.
OR
•
Click
Install Plug
to indicate that the gel pump plug (or empty cartridge) was
installed.
900493e.AI
900494e.AI
Cartridge Barrel
Cartridge Locking
Lever (open)
Cartridge
Wings
Gel Cartridge OR
Gel Pump Plug
900494e.AI
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...