Figure 149:New Binning Analysis window - Bin Parameters
A minimum of 10 samples is required to begin a Bin Analysis.
Bin Analysis Parameters
The first step in creating a new bin analysis is to set the bin analysis parameters. These
parameters identify which data in the fragment list will be included in the bin analysis
and also define the appearance of the view generated. Once the bin analysis
parameters have been set, the information from the selected result set will be used to
generate a scatter plot. This plot consists of all of the analyzed fragment data displayed
as a function of the peak height (y-axis) and fragment length (x-axis).
1. Select the
dye
that was used for the analysis of the raw data. Only one dye can be
examined at a time during each bin analysis.
2. Input the
fragment range
(the minimum and maximum fragment size).
3. Input the
maximum bin width
(maximum difference between the smallest and
largest fragments within a bin) in nucleotides.
4. Input the
minimum data points per bin
(the limiting number of points
necessary to form a bin in the scatter plot).
5. Input the
length
of the
repeat unit
of the allele.
6. Select the preferred
allele naming convention
(see the “Locus Tag Editor”
section for more information).
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...