Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-49
:BATTery[n]:SIMulator:VOC:EMPTy
This command sets the threshold Voc value for the discharging to stop in the battery simulator function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
Minimum
Usage
:BATTery[n]:SIMulator:VOC:EMPTy <value>
:BATTery[n]:SIMulator:VOC:EMPTy?
:BATTery[n]:SIMulator:VOC:EMPTy? DEFault
:BATTery[n]:SIMulator:VOC:EMPTy? MINimum
:BATTery[n]:SIMulator:VOC:EMPTy? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
The value range of the Voc: Minimum to Maximum
Details
This command sets the threshold Voc value for the discharging to stop in the battery simulator
function.
If the empty Voc is greater than the full Voc, the battery model cannot be accessed and you will see
error 703, "Full Voc can’t be less than empty Voc."
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.
Example
:BATT:SIM:VOC:EMPT 0.5
Set the empty Voc value to 0.5 V.
Also see
(on page 7-50)