:BATTery[n]:SIMulator:CURRent?
This command queries the real-time battery current of the battery simulator.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:SIMulator:CURRent?
[n]
Channel number: For single-channel instruments, the value is 1.
Example
:BATT:SIM:CURR?
The real-time battery current is displayed.
Also see
None
:BATTery[n]:SIMulator:CURRent:LIMit
This command sets the value of the current limit for stopping the discharging in the battery simulation function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0.1 A
Usage
:BATTery[n]:SIMulator:CURRent:LIMit <value>
:BATTery[n]:SIMulator:CURRent:LIMit?
:BATTery[n]:SIMulator:CURRent:LIMit? DEFault
:BATTery[n]:SIMulator:CURRent:LIMit? MINimum
:BATTery[n]:SIMulator:CURRent:LIMit? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 6.1 A
Details
This command sets the value of the current limit for stopping the discharging in the battery simulation
function.
This value cannot be changed when the simulation is in progress. If you modify this value without
pausing the simulation, error 704, "Not permitted while battery model is running" occurs.