Appendix B: Verification
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
B-16
077114601 / March 2019
Required equipment
These procedures use external equipment to directly check warranted characteristics. The following
table lists the required equipment.
Equipment
Required specifications
Recommended model
Connectors
Connect wire assemblies and test
resistors.
Linear AC power source
Variable AC output from 90 VAC to
265 VAC with at least 750 VA
capacity.
Electronic load
Variable DC load capable of 6 A
and 75 VDC
DC voltmeter
7½-digit high-accuracy voltmeter
Keithley 2001/2002
High current hook up wire
18 AWG minimum connection
wires
Low current hook up wire
22 AWG minimum connection
wires
50 m precision shunt resistor
50 mΩ precision shunt with 5 ppm
at 25 °C ±4 ppm/°C temperature
coefficient
Isotek
Corp RUG-Z-R050-0.1-TK1
5 precision shunt resistor
5 Ω precision shunt with 25 ppm at
25 °C ±4 ppm/°C temperature
coefficient
Isotek
Corp RUG-Z-5R00-0.1-TK1
Current sense resistor cabling
Kelvin 4 terminal measuring cables
from shunt resistor to voltmeter
150 resistor
Power rating > 5 W
Check DC voltage setting readback accuracy with remote sense
Equipment required
AC power source
High-current 18 AWG wire
Electronic load
Low-current 22 AWG wire
DC voltmeter
0.3 Ω resistors
To check DC voltage setting readback accuracy with remote sense:
1. Power off the device under test (DUT).
2. Remove the shorting jumpers that connect the output terminals to the sense terminals from the
output mating connector on the rear panel.