Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-137
[:SOURce[n]]:LIST:END:ZERO
This command specifies the source values when the instrument is stopping a list.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 (OFF)
Usage
[:SOURce[n]]:LIST:END:ZERO <value>
[:SOURce[n]]:LIST:END:ZERO?
[n]
Channel number: For single-channel instruments, the value is 1.
<value>
When the list is stopped,
Use the last source values:
OFF
or
0
Set the voltage to 0 V, current source to 0.1 A:
ON
or
1
Details
This command allows you to end a specified list with the last source value or zero value.
Example
LIST:END:ZERO ON
List ends with the 0 V and 0.1000 A.
Also see
None
[:SOURce[n]]:LIST:HTIMe
This command determines the hold time of the list running.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
POINt
Usage
[:SOURce[n]]:LIST:HTIMe <type>
[:SOURce[n]]:LIST:HTIMe?
[n]
Channel number: For single-channel instruments, the value is 1
<type>
POINt
: The list runs until the reading reaches the preset sample count; the hold time
is computed by sample count * sample speed; the preset dwell time is ignored in
this mode
TIME
: The list runs until the preset time is reached; the preset sample count is
ignored in this mode
Hold time is decided by whichever is longer: dwell time or sample count.