Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-62
077114601 / March 2019
:BATTery[n]:TEST[:SOURce]:LV
This command sets the low voltage (LV) value in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0 V
Usage
:BATTery[n]:TEST:[SOURce]:LV <value>
:BATTery[n]:TEST:[SOURce]:LV?
:BATTery[n]:TEST:[SOURce]:LV? DEFault
:BATTery[n]:TEST:[SOURce]:LV? MINimum
:BATTery[n]:TEST:[SOURce]:LV? MAXimum
[n]
Channel number: For single-channel instruments, the value is 1
<value>
0 to 21 V
Details
When the output voltage reaches this value, the digital I/O pin, which is set as Digital Out and LV
action, outputs a level signal.
Example
:BATT:TEST:LV 0.1
Set the low voltage as 0.1 V