Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 3: Functions and features
077114601 / March 2019
3-11
Data buffer
The data buffer captures measurements, instrument status, and the output state of the instrument.
The 2281S has only one buffer.
The data buffer provides the following statistics:
Average
Minimum
Maximum
Peak-to-peak
Standard deviation
You can perform the following operations on the data buffer from the front panel or the remote
interface:
Configure, store, and recall a reading buffer.
View reading buffer content.
Save reading buffer content to a USB flash drive.
Set reading buffer to next, always, or never fill mode.
Change the capacity of a reading buffer.
Clear a reading buffer manually or automatically.
Effects of reset and power cycle on buffer
The instrument clears the reading buffer when the instrument power is turned off and then on again.