Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Appendix B: Verification
077114601 / March 2019
B-43
Figure 146: 20 MHz voltage and current noise test setup
5. Set the oscilloscope as follows:
1 mV/division
1 MΩ input resistance
20 MHz bandwidth limit
AC coupled
Autotrigger
1 ms/div
Set to measure V
p-p
and V
RMS
6. Set the DUT to the 100% full-scale output voltage.
7. Set the DUT to the 100% full-scale output current.
8. Turn the DUT output on.
9. Enter the oscilloscope measurements in the
(on page B-14) table for