Section 2: General operation
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
2-20
077114601 / March 2019
Source Settings menu for the power supply function
You can change the following settings by pressing the
MENU
key, moving focus to the Source
Settings icon, and pressing the
Enter
key.
Setting
Description
V-Set
Configure the voltage setting using the numerical keys, soft keys, and
navigation control.
I-Limit
Configure the current setting using the numerical keys, soft keys, and
navigation control.
Output
Turn on or off the output. You can also disable the output. Once output is
disabled, the output is off and the OUTPUT key is ignored.
Output Delay
Enable or disable output delay.
Slew Rise
Set the rising rate for voltage source.
Slew Fall
Set the falling rate for voltage source.
Delay Rise
This menu option is only available if Output Delay is set to
On
. View the
delay time that the instrument waits before turning on the output.
Delay Fall
This menu option is only available if Output Delay is set to
On
. View the
delay time that the instrument waits before turning off the output.
Source Protection menu for the power supply function
This menu allows you to configure the protection settings.
Setting
Description
Overvoltage Protection
Set the overvoltage protection limit of the source output to restrict the
maximum voltage level that the instrument can source.
Overcurrent Protection
Set the overcurrent protection limit of the source output to restrict the
maximum current level that the instrument can source.
Maximum Voltage
Set the maximum voltage that can be set. This influences the voltage setting
range.