Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-175
:SYSTem:PRESet
This command returns the instrument to the factory default conditions and erases all user data stored in
nonvolatile memory.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:PRESet
Details
Sending this command will return the instrument to the factory default conditions and erases all user
data stored in nonvolatile memory, such as list parameter and setup information.
This command does not influence the status structure registers and error queue.
This command resets GPIB address and IP address to default settings. After sending this command,
make sure to use the new address to communicate with the instrument.
You should allow ample time for the command to be processed before attempting to communicate
with the instrument again.
Example
:SYST:PRES
Set the instrument to restore the factory default settings
and erase all user data stored in nonvolatile memory,
such as list parameter and setup information.
Also see
(on page 7-7)